Frequently Asked Questions
Got a question? There’s a good chance we’ve answered it here. Browse our FAQs or use the filters below to find what you're looking for. Couldn't find an answer to your question here? Contact us!

Find your question
Can't find an answer to your question? Contact us!
Search question
Product Category
Product
Can I re-use FabRICATOR Z Immobilized columns?
We can only guarantee optimal digestion for one-time use. We do not have a cleaning or regeneration protocol to provide. However, depending on the antibody and the following application the column can be reused. We do not recommend using with different antibodies due to the risk for contamination of carry-over from previous sample. If column reuse is desired store the column in 10-20% ethanol at +4-8°C.
Learn more about FabRICATOR®, FabRICATOR® Z
Can I release Fc fragments from the CaptureSelect column after isolating the F(ab’)2 fragments?
It is possible to elute the Fc fragments from the CaptureSelect column using a low pH buffer, for example a 0.1M Glycine buffer pH=3.0. The pH must immediately after elution be adjusted to neutral pH by adding 0.1x elution volume of 1M Tris pH=8.0.
Learn more about FabRICATOR®
Can I use protein A or MabSelectSure (GE) for removal of Fc-fragments?
Genovis has chosen CaptureSelect in the FabRICATOR Fab2 Kit for removal of Fc fragments. CaptureSelect has a very high specific affinity for the Fc. Both protein A and MabSelect Sure (GE) may have affinity for the Fab which may result in a lower recovery of the generated F(ab’)2 fragments and the degree of affinity is antibody dependent. We have many customer reports on this subject, and they usually come back to us for the CaptureSelect columns for the above reason.
Learn more about FabRICATOR®
Does FabRICATOR introduce modifications?
FabRICATOR is a cysteine protease. It should not result in any other modifications than the digestion of the peptide bond. Digestion will be between the two Glycines (…CPAPELLG / GPSVF…) and the C-terminus will regain the carboxyl group – that is addition of one OH-group and the N-terminal amino group will be regained (addition of one H).
In summary, R1-CO-NH-R2 + H2O –> R1-COOH + NH2-R2.
Learn more about FabRICATOR®
Can I use FabRICATOR Fab2 Kit for generation and purification of fragments from Rabbit IgG?
FabRICATOR Fab2 Kit digests rabbit IgG but to capture the Fc fragments in the affinity step you need another resin than the CaptureSelect™ Fc normally included in the kit. Please contact the Genovis team if you are interested in processing rabbit IgG and we will provide columns with resin compatible with rabbit Fc in the FabRICATOR Fab2 Kit.
Learn more about FabRICATOR®
What is the FabRICATOR MagIC?
The FabRICATOR MagIC magnetic beads with immobilized FabRICATOR (IdeS) enzyme.
Learn more about FabRICATOR®
Can I use FabRICATOR MagIC in a automation system?
FabRICATOR MagIC is suitable for using in automated systems that can handle magnetic beads. On the applications page you can read about how it’s used in a KingFisher™ workflow.
Learn more about FabRICATOR®, FabRICATOR® Xtra, FabRICATOR® Z
Which digestion buffers are compatible with FabRICATOR MagIC?
The protocol for FabRICATOR MagIC is based on using standard PBS pH 7.4 as digestion buffer. FabRICATOR in solution is compatible with most commonly used buffers at pH=6-8 and the same buffers should work for FabRICATOR MagIC, but the incubation time might need to be optimized.
Learn more about FabRICATOR®
