Can I use protein A or MabSelectSure (GE) for removal of Fc-fragments?
Genovis has chosen CaptureSelect in the FabRICATOR Fab2 Kit for removal of Fc fragments. CaptureSelect has a very high specific affinity for the Fc. Both protein A and MabSelect Sure (GE) may have affinity for the Fab which may result in a lower recovery of the generated F(ab’)2 fragments and the degree of affinity is antibody dependent. We have many customer reports on this subject, and they usually come back to us for the CaptureSelect columns for the above reason.
Learn more about FabRICATOR®
Can I release Fc fragments from the CaptureSelect column after isolating the F(ab’)2 fragments?
It is possible to elute the Fc fragments from the CaptureSelect column using a low pH buffer, for example a 0.1M Glycine buffer pH=3.0. The pH must immediately after elution be adjusted to neutral pH by adding 0.1x elution volume of 1M Tris pH=8.0.
Learn more about FabRICATOR®
Which digestion buffers are compatible with FabRICATOR HPLC?
We have tested PBS and 150 mM ammonium acetate, pH7 with good results. More buffers were found to be compatible with FabRICATOR digestion in solution (more info) and should be suitable for FabRICATOR HPLC as well. However, the ionic strength of the buffer influences retention of the antibody fragments on the FabRICATOR HPLC column and a minimum of 100-150 mM salt is necessary for IgG1 subunits to elute as a single, narrow peak. Other IgG subclasses or fusion proteins might need some optimization of buffer composition.
Learn more about FabRICATOR®
Can I re-use FabRICATOR Immobilized columns?
We can only guarantee optimal digestion for one-time use. We do not have a cleaning or regeneration protocol to provide. However, depending on the antibody and the following application the column can be reused. We do not recommend using with different antibodies due to the risk for contamination of carry-over from previous sample. If column reuse is desired store the column in 10-20% ethanol at +4-8°C.
Learn more about FabRICATOR®
What is the FabRICATOR MagIC?
The FabRICATOR MagIC magnetic beads with immobilized FabRICATOR (IdeS) enzyme.
Learn more about FabRICATOR®
Which digestion buffers are compatible with FabRICATOR MagIC?
The protocol for FabRICATOR MagIC is based on using standard PBS pH 7.4 as digestion buffer. FabRICATOR in solution is compatible with most commonly used buffers at pH=6-8 and the same buffers should work for FabRICATOR MagIC, but the incubation time might need to be optimized.
Learn more about FabRICATOR®
Can I use FabRICATOR MagIC in a automation system?
FabRICATOR MagIC is suitable for using in automated systems that can handle magnetic beads. On the applications page you can read about how it’s used in a KingFisher™ workflow.
Learn more about FabRICATOR®
Can I use FabRICATOR and PNGaseF in the same reaction?
Yes, PNGaseF and FabRICATOR can be used in a one-pot reaction as long as the reaction conditions follow the specifications for the FabRICATOR enzyme. It is important to have native reaction conditions for FabRICATOR to remain active.
Learn more about FabRICATOR®
Can I use FabRICATOR Fab2 Kit for generation and purification of fragments from Rabbit IgG?
FabRICATOR Fab2 Kit digests rabbit IgG but to capture the Fc fragments in the affinity step you need another resin than the CaptureSelect™ Fc normally included in the kit. Please contact the Genovis team if you are interested in processing rabbit IgG and we will provide columns with resin compatible with rabbit Fc in the FabRICATOR Fab2 Kit.
Learn more about FabRICATOR®
Can FabALACTICA and FabRICATOR be used in the same reaction?
Yes, for human IgG1 molecules both enzymes can be used in combination to generate a specific hinge peptide. The incubation time is significantly longer for FabALACTICA and it is recommended to first run the FabALACTICA digest and then add FabRICATOR for the last 30 min.
Learn more about FabALACTICA™, FabRICATOR®